• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC T2 STRATOSPHERE EXPANDABLE CORPECTOMY SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC T2 STRATOSPHERE EXPANDABLE CORPECTOMY SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE Back to Search Results
Model Number 436113C
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare professional via manufacturer representative regarding a patient having degenerative cervical stenosis for spinal therapy.It was reported that cage was correctly loaded on the holder but surgeon was not able to expand it while turning the t-handle.The product was implanted on (b)(6) 2020 and was explanted on the same day.There was 30 mins of procedural delay.There were no patient symptoms/ complications as a result of the event.
 
Additional Manufacturer Narrative
H3: product analysis; part # 436113c; lot # ca19c042 visual and optical inspection of the centerpiece expander did not reveal any damages that would hinder the implant from expanding.Functional inspection with a sample 13mm inserter confirmed the implant was able to connect and engage, expand and close without any grinding or restrictions.The implant appears to function as intended.No fault found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T2 STRATOSPHERE EXPANDABLE CORPECTOMY SYSTEM
Common Device Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key11153215
Report Number1030489-2021-00046
Device Sequence Number3756113
Product Code MQP
UDI-Device Identifier00643169634817
UDI-Public00643169634817
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K183510
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number436113C
Device Catalogue Number436113C
Device Lot NumberCA19C042
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2020
Supplement Date Received by Manufacturer04/01/2021
Initial Report FDA Received Date01/12/2021
Supplement Report FDA Received Date04/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-