Model Number 7540220
Medical Device Problem Codes Device Slipped (1584); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare professional via field representative regarding patient with unknown indication involved in posterior thoracolumbar surgery.It was reported that during procedure, when the screw plug was tightened, the tip of the tool driver was broken in the screw plug, making the screw plug unusable.In the process of locking the crosslink, the screw plug was tightened, and the screw plug was slipped.As a result, crosslink could not be locked and it could not be installed.Reported set screw was not implanted.No fragments left in the patient's body.Event was associated with the patient.
 
Additional Manufacturer Narrative
H3: device evaluated summary - visual and optical inspection revealed that the hex of the screw appears to be stripped.This is consistent with overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key11153769
Report Number1030489-2021-00047
Device Sequence Number3756114
Product Code KWP
UDI-Device Identifier00885074468174
UDI-Public00885074468174
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K052187
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7540220
Device Catalogue Number7540220
Device Lot NumberH5495654
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2020
Supplement Date Received by Manufacturer05/12/2021
Initial Report FDA Received Date01/12/2021
Supplement Report FDA Received Date05/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-