| Model Number |
407201 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Great Vessel Perforation (2152)
|
| Date of Event |
12/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported perforation could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
|
| |
|
Event or Problem Description
|
|
During an atrial tachycardia ablation procedure for the left atrium, while conducting the transseptal puncture, a perforation was noted on one of the leaflets of the aortic valve.After performing the transseptal puncture, when passing the guidewire through, the guidewire went down into the aorta.The perforation was diagnosed initially with ultrasound and then later via intracardiac echocardiogram.The procedure was then cancelled and no intervention was administered.There were no performance issues with the abbott product.
|
| |
|