• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P FMP EXTENDED LINERS; LINER/NON-HOODED-NEU, MP8, HXE-PLUS, 40MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P FMP EXTENDED LINERS; LINER/NON-HOODED-NEU, MP8, HXE-PLUS, 40MM Back to Search Results
Model Number 931-40-754
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/15/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - exchanged head and liner due to infection.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 26 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was a nonconformance associated with the main part #931-40-754, liner/non-hooded-neu, mp8, hxe-plus, 40mm, which documents that out of 10 parts lot, 1 part was scrapped due to machine moved during marking the component.All other item in the lot met with the design, fit and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post-surgical instructions.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP EXTENDED LINERS
Common Device Name
LINER/NON-HOODED-NEU, MP8, HXE-PLUS, 40MM
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin,tx 78758-5445
MDR Report Key11155574
Report Number1644408-2020-01241
Device Sequence Number10954328
Product Code OQG
UDI-Device Identifier00190446177920
UDI-Public(01)00190446177920
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K172651
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/24/2025
Device Model Number931-40-754
Device Catalogue Number931-40-754
Device Lot Number011V1105
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/15/2020
Supplement Date Received by Manufacturer03/09/2021
Initial Report FDA Received Date01/12/2021
Supplement Report FDA Received Date04/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-402 LOT 871B1260; 400-03-402 LOT 871B1260
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
-
-