MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC ESPRIT VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC ESPRIT VENTILATOR Back to Search Results
Model Number V1000
Device Problems Defective Device (2588); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 12jan2021.
 
Event Description
The biomed reported to philip's remote service engineer (rse) hearing a hissing noise inside the unit by the oxygen (o2) regulator.The biomed advised that staff heard noise when the unit was in the storage room.The rse reached out to senior tech who advised the oxygen regulator is defective.The rse provided part identification for oxygen regulator assembly.The device did not have patient involvement at the time the issue was discovered, therefore, there was no patient or user harm reported.The biomed replaced oxygen regulator assembly to resolve the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESPRIT VENTILATOR
Type of Device
VENTILATOR
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
carlsbad CA
Manufacturer Contact
bill cole
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key11155700
MDR Text Key227750974
Report Number2031642-2021-00120
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K981072
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV1000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/15/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse