| Model Number |
1012452-15 |
| Medical Device Problem Codes |
Material Separation (1562); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat an unspecified lesion.A 3.75x15mm nc trek balloon was advanced over the guide wire when it became kinked and subsequently the mid-shaft broke (having not entered the anatomy yet).The device was simply removed altogether.Another nc trek balloon was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed on the returned device.The reported separation and kink were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulties appear to be related to operational context.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|