• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. ARROW VPS RHYTHM MONITOR BUNDLE; CATHETER INTRAVASCULAR THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. ARROW VPS RHYTHM MONITOR BUNDLE; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Model Number IPN037371
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2020
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).
 
Event Description
Customer reports "unreliable / inaccurate picc tip positioning's using the ecg.The catheter to be adjusted post insertion by at least 3 cm i.E.The ecg placed the lines short even though we had what looked like a satisfactory ecg." no patient harm reported.
 
Event Description
Customer reports "unreliable / inaccurate picc tip positioning's using the ecg.The catheter to be adjusted post insertion by at least 3cm i.E.The ecg placed the lines short even though we had what looked like a satisfactory ecg." no patient harm reported.
 
Manufacturer Narrative
Qn#(b)(4).The report there was a correlation issue between the internal ecg waveform and the confirming xray was not confirmed since the sample was not returned for review.A photograph of some snapshots of the ecg signal was provided by the complainant.The event details and photo were reviewed.It is possible that they were threading very quickly and there was a delay in the readings on the screen; although that could not be confirmed without other evidence.It should be noted, the vps rhythm device does not record the entire picc placement procedure therefore a review of the case is not possible.A device history record review was performed and did not reveal any manufacturing related issues.A probable cause of this issue could not be determined based on the information provided and without a sample.No further action will be taken.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW VPS RHYTHM MONITOR BUNDLE
Type of Device
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key11156379
MDR Text Key226329635
Report Number3011137372-2021-00010
Device Sequence Number1
Product Code LJS
Combination Product (y/n)N
PMA/PMN Number
K160925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN037371
Device Catalogue NumberRHY-100-MTRB
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/21/2020
Initial Date FDA Received01/12/2021
Supplement Dates Manufacturer Received01/29/2021
Supplement Dates FDA Received02/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-