MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TIBIAL BROACH IMPACTOR; INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TIBIAL BROACH IMPACTOR; INSTRUMENT Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: foreign (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during a total knee arthroplasty the tibial broach handle fractured during the procedure.No pieces fell into the patient and the procedure was completed without delay.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).The complaint is confirmed.Visual examination of the provided pictures identified the head of the instrument was fractured.The picture shows signs of use.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g4; g6; h1; h2; h3; h6 complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product identified signs of repeated use (impaction marks, gouges, and nicks).The weld which connects the rod head with the rod shaft was fractured.The potential field age of this instrument is 4 years and 1 month.Device history record (dhr) was reviewed and no discrepancies were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL BROACH IMPACTOR
Common Device Name
INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11156753
Report Number0001822565-2021-00100
Device Sequence Number18993221
Product Code HWA
UDI-Device Identifier00889024207707
UDI-Public(01)00889024207707(10)63561564
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00595109000
Device Lot Number63561564
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/05/2021
Supplement Date Received by Manufacturer02/23/2021
05/03/2021
Initial Report FDA Received Date01/12/2021
Supplement Report FDA Received Date02/25/2021
05/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-