| Model Number |
364505 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was chosen for treatment of a severely calcified lesion (90 percent stenosis degree) in a mildly tortuous lad.It was not possible to push the stent across another stent on lm-plad.The stent was withdrawn and checked which showed that one strut of the stent was deformed and protruded.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the stent is severely deformed at its proximal end.Several stent struts are lifted.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the lifted struts were initially crimped on the balloon.The balloon is still well folded and shows no signs of inflation.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
|
| |
|