Model Number N600X-1
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code Diminished Pulse Pressure (2606)
Date of Event 11/05/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During home monitoring use the n600x-1 spo2 pulse rate alarm did not go off at the time of patient¿s death.The patient was paralyzed from the neck down, ventilator dependent with a pacemaker and peg tube for feedings.The pacemaker was 10 years old.The patient caregivers informed the police that the device did not alarm at the time of death, and that another monitor (mindray) was being used at the same time to monitor the heartbeat which was decreasing.It was reported that the n600x-1 was reading 60bpm, which was what the pacemaker was set to.Following the event, nhs trust performed 24-hour testing of the n600x-1 and simulated the reduction of pulse and saturation and the device alarmed as normal.Per the reporter the patients cause of death was pneumonia.The device it not related to the cause of death.
 
Additional Manufacturer Narrative
Correction: b2, h1 additional information: a2 (age at event, date of birth).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the battery had failed as the device was an aftermarket battery not approved by medtronic and was expired.It was reported that the monitor displayed higher readings than expected.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
OXIMETER
Manufacturer (Section D)
NELLCOR PURITAN BENNETT IRELAND
micheal collins rd mervue
galway
EI 
MDR Report Key11157932
Report Number8020893-2021-00012
Device Sequence Number3583680
Product Code DQA
UDI-Device Identifier10884521181229
UDI-Public10884521181229
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K123581
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN600X-1
Device Catalogue NumberN600X-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2021
Initial Date Received by Manufacturer 12/14/2020
Supplement Date Received by Manufacturer01/12/2021
07/13/2021
Initial Report FDA Received Date01/12/2021
Supplement Report FDA Received Date02/08/2021
07/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age57 YR
-
-