| Model Number |
1011493-18 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly calcified de novo renal artery that is 70% stenosed.A 5.0x18mm rx herculink elite stent delivery system completely failed to deploy.Device was removed and another same size herculink elite was used to successfully complete the procedure.There were no adverse patient effects reported and no clinically significant delay.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported difficulty to deploy was not confirmed.The noted stretch on the guidewire notch is likely due to case circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.A conclusive cause for the reported deployment failure was unable to be determined as the issue was not confirmed on the returned unit.It may be possible that the inflation device being used was faulty or not properly connected; however, this could not be confirmed.The stretched guidewire notch is likely due to inadvertent mishandling.This damage can occur if an attempt is made to pull the stent delivery system in an opposite direction as the guide wire.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|