OWENS & MINOR IHALYARD MANUAL SURGICAL INSTRUMENT FOR GENERAL USE; INST: SCALPEL METAL NUM 10 20/BX Back to Search Results
Model Number 56214
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Injury (2348)
Date of Event 12/15/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.Upon completion of the investigation; a follow-up report will be filed.All information reasonably known as of 12jan2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by owens & minor, inc.Represents all of the known information at this time.Owens & minor, has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the owens & minor complaint database and identified as complaint (b)(4).
 
Event or Problem Description
Mai complaint number - (b)(4).Before performing a procedure on a patient, the health care provider had to use extreme excessive force to remove the plastic covering of blade and was cut in the left hand when trying to remove this covering.Per the customer, "provider will need 48 hour wound check, after receiving # 5 sutures (b)(6) 2020, day of receiving laceration injury from described incident, above, and said provider will need sutures removed 9-11 days after sutures applied.Provider made incident report to (b)(6) on day of injury, (b)(6) 2020.The scalpel that i was cut by, after excessive force removing the "point protector", is on the left.The plastic cover is 3 notches down, covering the handle, versus the other scalpel is not down as far, and just covering handle flat area." no patient involvement.Hibc/gtin of product: (b)(4).
 
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.The photos of the kit labeling for both complaint lots, as well as a close-up photo of two unopened packages compared two scalpels, one showed the point protector farther down on the handle than the other scalpel.At the time of complaint receipt, inventory of the complaint kit lots and scalpel lots involved were depleted.A statistical sampling was performed on the current scalpel volume which is a different lot than the reported lot.All samples inspected met product specifications.The point protector was placed appropriately, and excessive force was not needed to remove the covering.A five-year complaint history of the kit (56214) as well as the kit lots associated with the complaint incident did not indicate any additional reports of an injury involving the scalpel or issues with the point protector.Based on our investigation and trending analysis, this complaint does not appear to be a widespread issue but isolated in nature.Actions taken: as part of our investigation process, a supplier corrective action request (scar) was issued to our scalpel supplier.If samples are received at a later date, we will forward to our supplier for examination.This complaint has been logged in our complaint database where we will continue to monitor for any related trends.All information reasonably known as of 29jan2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by owens & minor represents all of the known information at this time.Owens & minor has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the owens & minor complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a owens & minor product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MANUAL SURGICAL INSTRUMENT FOR GENERAL USE
Common Device Name
INST: SCALPEL METAL NUM 10 20/BX
Manufacturer (Section D)
OWENS & MINOR IHALYARD
one edison drive
alpharetta GA 30005
MDR Report Key11158175
Report Number1030451-2021-00001
Device Sequence Number14274231
Product Code KDC
UDI-Device Identifier20809160003760
UDI-Public20809160003760
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date12/14/2022
Device Model Number56214
Device Catalogue Number56214
Device Lot Number240862
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/13/2021
Supplement Report FDA Received Date02/01/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-