| Model Number |
N/A |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information: unknown.The issue was evaluated and replicated in the fmsu lab; the cause was traced to a software configuration.This issue is very rare and requires many tries with large studies to reproduce; it is considered to be a very rare occurrence in a clinical environment.The numeric id assigned to a patient record could be reused once a newer patient record was merged to an older patient record.When this action is done, the newer id is made available again.This was not an issue prior to advanced reporting.With the introduction of advanced reporting the numeric id is not released within the advanced reporting database schema and that is how the mismatch between the patient demographic data can occur.Ref: internal complaint number comp (b)(4).
|
| |
|
Event or Problem Description
|
|
On december 02, 2020 fujifilm medical systems u.S.A., inc.(fmsu) service department received a customer inquiry for assistance with synapse cardiovascular.Study was sent in for a new patient with the correct mrn.Incorrect mrn is showing in cv client.On december 16, 2020, a risk assessment was performed to investigate the risk to patient safety.There was no patient impact, serious injury, or death associated with this event.The issue is considered highly detectable by a healthcare professional.
|
| |
|
Additional Manufacturer Narrative
|
|
Fujifilm initiated a recall on (b)(6) 2021 to alert customers of the patient id issues existing in synapse cv 6 users with advanced reporting.Fujifilm submitted c&r report (1000513161-05/05/2021-001-c) to the fda, which was classified as class ii and assigned recall number z-1775-2021 on 06/02/2021.No further investigation is necessary.
|
| |
|
Search Alerts/Recalls
|