Model Number TV-AB2680-J
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/18/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device remains implanted.
 
Event or Problem Description
The patient was initially implanted with the ovation ix abdominal stent graft system to treat an abdominal aortic aneurysm (aaa).During this initial a procedure, a possible type 1a endoleak was observed.The physician elected to implant a palmaz (non- endologix) stent proactively.This event was previously reported under (mdr # 300811247-2019-00110).Now, approximately 1.5 years post initial procedure, a type 1a endoleak was observed during routine follow up.Patient is currently being monitored while an intervention is being discussed.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type 1a endoleak as unconfirmed.This is not consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy relatedness of this complaint could not be determined with the medical records available for review.The final patient status was reported as pending re-intervention.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g1,2: contact office- name has been updated.G4: date of received by manufacturer.H6: result code: remove code 3233.H6: conclusion code: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Common Device Name
MAIN BODY
Manufacturer (Section D)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
MDR Report Key11159014
Report Number3008011247-2021-00005
Device Sequence Number19734659
Product Code MIH
UDI-Device IdentifierM701TVAB2680J1
UDI-Public+M701TVAB2680J1/$$3220419FS041219066
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/19/2022
Device Model NumberTV-AB2680-J
Device Catalogue NumberTV-AB2680-J
Device Lot NumberFS041219-06
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/18/2020
Supplement Date Received by Manufacturer02/22/2021
Initial Report FDA Received Date01/13/2021
Supplement Report FDA Received Date03/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OVATION IX ILIAC LIMB, LOT # FS071018-30; OVATION IX ILIAC LIMB, LOT # FS072618-02; OVATION PRIME FILL POLYMER, LOT # FF112918-01; PALMAZ (NON- ENDOLOGIX) STENT; OVATION IX ILIAC LIMB, LOT # FS071018-30; OVATION IX ILIAC LIMB, LOT # FS072618-02; OVATION PRIME FILL POLYMER, LOT # FF112918-01; PALMAZ (NON- ENDOLOGIX) STENT
Outcome Attributed to Adverse Event Other;
Patient Age89 YR
-
-