MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number TMICL13.2
Medical Device Problem Codes Patient-Device Incompatibility (2682); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation type 4110: lens work order search-no similar complaint type events reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2 implantable collamer lens, -16.0/+2.5/089 (sphere/cylinder/axis), into the patient's eye on (b)(6) 2020.Reportedly, the lens kept rotating and was explanted on (b)(6) 2020.It was not replaced due to patient anatomy.No patient injury reported.
 
Additional Manufacturer Narrative
B5-additional information: apparently this occurred due to the patient's anatomy.During the surgery the patient's eye wouldn't adhere to the drops in order to place the icl into the patient's eyes.Claim# (b)(6).
 
Additional Manufacturer Narrative
B5-additional information: the lens was implanted into the patient's left (os) eye.H3-device evaluation: the lens was returned in liquid in a specimen cup.Visual inspection found that the haptic was torn.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key11159186
Report Number2023826-2021-00116
Device Sequence Number10971956
Product Code QCB
UDI-Device Identifier00841542109888
UDI-Public00841542109888
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model NumberTMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2021
Initial Date Received by Manufacturer 12/22/2020
Supplement Date Received by Manufacturer01/13/2021
02/23/2021
Initial Report FDA Received Date01/13/2021
Supplement Report FDA Received Date02/08/2021
03/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
-
-