| Model Number |
N/A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Implanted date: device was not implanted.Explanted date: device was not explanted the actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
|
| |
|
Event or Problem Description
|
|
The user facility reported a kink in the destination sheath.The sheath was used for treatment of severe calcified lesion in the superficial femoral artery (sfa).There were severe calcification and tortuosity of the vessel, a kink of the proximal part of the sheath was confirmed.The sheath was not used and was replaced with another sheath to continue the procedure.The procedure was successfully completed.There was no health damage for the patient.The estimated blood loss was less than 250cc.There was no patient injury, medical/surgical intervention required.There were no other devices or equipment used with the reported product.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed for mechanical damage since the sample was not returned for evaluation.Based on the information given, the exact root cause of the event cannot be determined.There is no indication that any manufacturing, design or quality system issues may have led to this event.Currently, no action is recommended since this risk evaluation is within the predetermined limits in the fmea.
|
| |
|
Search Alerts/Recalls
|