ENCORE MEDICAL L.P X-ALT HIP; LINER 10 DEG HOODED-NEU, MP9, HXE-PLUS, 36MM Back to Search Results
Model Number 932-36-756
Health Effect - Clinical Code Fall (1848)
Date of Event 12/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - patient fell days after initial surgery.Hip was dislocating, so a constraining liner was needed.
 
Additional Manufacturer Narrative
Corrected data: see g.1.Manufacturer narrative: the reason for this revision surgery was reported as dislocation after a fall.The previous surgery and the surgery detailed in this event occurred 30 days apart.The healthcare professional indicated that there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocation after a fall.There were no findings during this evaluation that indicate the reported devices were defective.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-ALT HIP
Common Device Name
LINER 10 DEG HOODED-NEU, MP9, HXE-PLUS, 36MM
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin,tx 78758-5445
MDR Report Key11163278
Report Number1644408-2020-01270
Device Sequence Number3285832
Product Code OQG
UDI-Device Identifier00888912100182
UDI-Public(01)00888912100182
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130365
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/03/2025
Device Model Number932-36-756
Device Catalogue Number932-36-756
Device Lot Number904N2021
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/16/2020
Supplement Date Received by Manufacturer02/24/2021
Initial Report FDA Received Date01/13/2021
Supplement Report FDA Received Date03/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-364 LOT 869B1169; 400-03-364 LOT 869B1169
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight98
-
-