• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P EXPRT KNEE; INSERT, REVISION EXPRT, SZ B X17MM, E-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P EXPRT KNEE; INSERT, REVISION EXPRT, SZ B X17MM, E-PLUS Back to Search Results
Model Number 315-0B-717
Health Effect - Clinical Codes Pain (1994); Balance Problems (4401)
Date of Event 12/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - patient was unstable and complaining of pain.Surgeon increased poly to a 19mm.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as instability and pain.The previous surgery and the surgery detailed in this event occurred 1 year and 11 months apart.Initial or prolonged hospitalization was required.The healthcare professional indicated that there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to instability and pain.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT KNEE
Common Device Name
INSERT, REVISION EXPRT, SZ B X17MM, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
MDR Report Key11163294
Report Number1644408-2020-01218
Device Sequence Number3285833
Product Code JWH
UDI-Device Identifier00888912114011
UDI-Public(01)00888912114011
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140830
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/08/2023
Device Model Number315-0B-717
Device Catalogue Number315-0B-717
Device Lot Number009R1017
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/16/2020
Supplement Date Received by Manufacturer03/02/2021
Initial Report FDA Received Date01/13/2021
Supplement Report FDA Received Date03/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age62 YR
-
-