The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection revealed that the metal shaft was broken near the handle, with minimal bending present.There was some discoloration on the handle, but minimal debris.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found the following warnings and precautions related to the reported failure: ¿read these instructions completely prior to use.For single use only.Discard any open, unused product.Do not use after the expiration date.It is the surgeon¿s responsibility to be familiar with the appropriate surgical techniques prior to use of this device.¿ the complaint was confirmed, but the root cause could not be established.Factors that could have contributed to the reported event include development of a hairline crack in transport, excessive force on the device, use of the device as a lever, or an inadvertent impact event in transport or at the customer site.No containment or corrective actions are recommended at this time.
|