• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG AS VEGA PS TIBIAL PLATEAU CEMENTED T1; KNEE ENDOPROSTHETICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG AS VEGA PS TIBIAL PLATEAU CEMENTED T1; KNEE ENDOPROSTHETICS Back to Search Results
Model Number NX051Z
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Implant Pain (4561); Swelling/ Edema (4577)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation results: the affected device was not available for investigation.Therefore an investigation of the device was not possible.Due to the lack of information the device quality and manufacturing history records (dhr) can not be checked.The determination of a definite root cause is due to the non-availability of important information and the device itself not possible.The conclusion from the root cause analysis is that a systematic device deviation is unlikely since the received complaints are limited to certain health care institutions and their applicants.The participation of the applicants in this deviation scenario cannot be excluded by today.Based on the root cause analysis result no corrective actions have been carried out.As a preventive action we were in contact with the affected health care institutions and applicants in order to reduce the probability of recurrence.The complaints were submitted to aesculap ag as a summary report.In order to examine further actions, all cases from the summary report are processed as vigilance suspicion cases and are continuously monitored.
 
Event or Problem Description
It was reported that there was an issue with nx051z-as vega ps tibial plateau cemented t1.It was reported on (b)(6) 2019 that as a result of having the product implanted, the patient has experienced pain and swelling in both knees, and normal daily activities were restricted due to this pain.Intraoperative findings during replacement surgeries were loosening of the tibial and femoral components which resulted in a revision surgery.The primary surgery occurred on (b)(6) 2013, and the revision surgery occurred on (b)(6) 2018.All available information has been provided at this time, if additional information becomes available the complaint will be updated accordingly.Cement used: palacos r radiopaque bone cement.Involved components: nx042 (universal patella p2), nx112 (ps pe insert t1/t1+, 14mm), nn260p (peek plug f/ tibia).The adverse event / malfunction is filed under aag reference 100026394/400485940.Associated medwatch report: 9610612-2020-00558 (nx011z 400485938), 9610612-2020-00559 (nx051z 400485940).The same patient has an adverse event / malfunction filed under reference (b)(4) ( right knee).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA PS TIBIAL PLATEAU CEMENTED T1
Common Device Name
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key11166030
Report Number9610612-2020-00559
Device Sequence Number19762443
Product Code JWH
UDI-Device Identifier04046963810375
UDI-Public4046963810375
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K101281
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNX051Z
Device Catalogue NumberNX051Z
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2019
Initial Report FDA Received Date01/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-