It was reported that, during a navio dry bone lab/demo, the anspach drill kept heating up so much that it was extremely hot to the touch.The error code present was an health professional.They swapped the drill for its backup to continue.No patient involved.No other complications were reported.
|
H3, h6: the navio drill, part number 101209, (b)(6) , intended for use in treatment was returned for evaluation.There was a relationship between the device and the reported events.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was confirmed.The drill was overheating.The root cause was determined to be component failure.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
|