Model Number 7655405
Device Problem Connection Problem (2900)
Patient Problem Insufficient Information (4580)
Event Date 12/15/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of redu3920 showed eight other similar product complaint(s) from this lot number.The complaints for this lot number have been reported from the same facility in (b)(4).
 
Event Description
It was reported that after the catheter was placed, the extension tubing was found unable to be firmly engaged.It was reported this occurred with nine devices.This report addresses the ninth device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GROSHONG NXT CLEARVUE 4F SL BASIC TRAY
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key11166726
MDR Text Key233890778
Report Number3006260740-2021-00026
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00801741035272
UDI-Public(01)00801741035272
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K034020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7655405
Device Catalogue Number7655405
Device Lot NumberREDU3920
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-