Model Number 1013470-200
Medical Device Problem Codes Material Rupture (1546); Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no indication of a lot specific product quality issue.Based on the information provided, the reported resistance during advancement and balloon rupture appear to be due to circumstances of the procedure.It is likely that the resistance and balloon rupture occurred due to interaction with the lesion calcification causing damage to the outer surface of the balloon resulting in rupture during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
 
Event or Problem Description
It was reported that the procedure was performed to treat a mildly calcified, mildly tortuous superficial femoral artery.The 6.0x200mm armada 18 balloon dilatation catheter was positioned with resistance during advancement due to the anatomy.During inflation, the balloon ruptured at 10atm on the first inflation.The balloon catheter was removed without issue and replaced with a new armada 18 to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11166818
Report Number2024168-2021-00415
Device Sequence Number3717576
Product Code LIT
UDI-Device Identifier08717648208461
UDI-Public08717648208461
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model Number1013470-200
Device Catalogue Number1013470-200
Device Lot Number9121341
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/06/2021
Initial Report FDA Received Date01/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Patient Weight105
-
-