• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR HI-TORQUE WHISPER MS GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR HI-TORQUE WHISPER MS GUIDE WIRE Back to Search Results
Model Number 1005357H
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Arrest (1762); Low Blood Pressure/ Hypotension (1914); Perforation (2001)
Date of Event 12/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
There was no reported device malfunction and the product was not returned.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of perforation is listed in the instructions for use for hi-torque guide wires, ce/fda as a known patient effect of the procedure.A conclusive cause for the reported patient effects and the relationship to the product, if any, cannot be determined.Additionally, the reported treatments appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the moderately calcified, moderately tortuous, 90% stenosed distal left anterior descending coronary artery.A whisper ms guide wire crossed the lesion, a 2x15mm trek balloon was used for pre-dilatation, and a 2.25x23mm xience sierra stent was implanted without issue.As an unspecified pressure wire was advanced, the whisper guide wire may have accidentally advanced and perforated the vessel, so medication was used as treatment.The patient's blood pressure decreased, so pericardiocentesis and an intra-aortic balloon pump were used.The patient went into cardiac arrest and expired that same day(b)(6) 2020.In the opinion of the physician, the use of these abbott devices did not cause or contribute to the patient death in any way.There was no clinically significant delay in the procedure.In the opinion of the physician, the use of these devices did not cause or contribute to complications or adverse events.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE WHISPER MS GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11167256
Report Number2024168-2021-00421
Device Sequence Number3718196
Product Code DQX
UDI-Device Identifier08717648037801
UDI-Public08717648037801
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K101116
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model Number1005357H
Device Catalogue Number1005357H
Device Lot Number0071671
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/22/2020
Initial Report FDA Received Date01/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient Weight55
-
-