| Model Number |
67P035 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the device cuff would not inflate.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h10 ? device evaluation results: the affected bivona tracheostomy tube was returned for investigation in used condition.The customer reported product problem (inflation failure) was not confirmed during functional testing.The cuff inflated as intended.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.No fault was found with the device.
|
| |
|