STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES, 50-S; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279351250
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the device was continuously activating and the device arced.
 
Event or Problem Description
It was reported that the device was continuously activating and the device arced.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: continuous activation.Probable root cause: probe short.Button stuck on firing position.Fluid ingress.Suction tube cut.Button inadvertently pressed.Damaged insulation.Probe bender used to bend nonbendable probe.User accidentally submerges device.Inadequate gluing operations or inadequate gluing/welding of core to handle.Console error.Use error.The reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RF 2 PROBES, 50-S
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11167535
Report Number0002936485-2021-00027
Device Sequence Number2766333
Product Code GEI
UDI-Device Identifier07613327056983
UDI-Public07613327056983
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K071859
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0279351250
Device Catalogue Number0279351250
Device Lot Number19102AE2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/15/2020
Supplement Date Received by Manufacturer12/15/2020
Initial Report FDA Received Date01/13/2021
Supplement Report FDA Received Date02/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-