Model Number MT20649-3 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/27/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that unexpected receiver shutdown occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
|
Event Description
|
It was reported that unexpected receiver shutdown occurred.The product was evaluated.An external visual inspection was performed and passed.Charge and boot test was performed and passed.Receiver functional test was performed and passed.An internal visual inspection was performed and passed.The log was downloaded and reviewed finding no errors related to the complaint.The allegation was not confirmed.The probable cause could not be determined post-investigation as the allegation was not found.No injury or medical intervention was reported.
|
|
Manufacturer Narrative
|
(b)(4).B5: describe event or problem additional.D9: device availability additional.H2: additional information/device evaluation.H3: device evaluated by manufacturer additional.H6: event problem and evaluation codes additional.
|
|