• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING USA, LLC WHITESTAR SIGNATURE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING USA, LLC WHITESTAR SIGNATURE; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number OPO73
Medical Device Problem Codes Break (1069); Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event is unknown as it was not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
A sterility assurance concern was reported.A brief description from the surgery center indicated that there was damage to the opo73 phaco tubing pack tray and they were unsure if the packaging had been comprised and the tubing pack was no longer sterile.
 
Additional Manufacturer Narrative
Section d10: device available for evaluation: yes.Section d10: returned to manufacturer on: 1/26/2021.Section h3: device returned to manufacturer: yes.Device evaluation: the sealed opo73 was received confirming the reported lot number 60242584.A visual inspection of the returned product revealed minor cosmetic damage to the universal tray.No breach of the sterile barrier was observed and the tyvek lid seal is still intact.The reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Additional Manufacturer Narrative
H3: a record review was performed.A product deficiency review was performed and there is no product deficiency identified.Device trending was reviewed.There is not a recognizable adverse trend.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.A labeling review was conducted and found to include adequate instructions for use, warnings and operational errors.The review of the device history record (dhr) for the opo73 fusion dual pump pack, lot #: 60242584 showed that there were no issues or non-conformities.The device and its components met all specifications prior to being released.Manufacturing has been ruled out as a potential cause for the reported issue.Based on the investigation results, no corrective action has been issued.Based on the investigation results there is no indication of a product quality deficiency.Johnson & johnson surgical vision will continue to monitor this type of complaint.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WHITESTAR SIGNATURE
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
MDR Report Key11168071
Report Number3006695864-2021-07040
Device Sequence Number19390955
Product Code HQC
UDI-Device Identifier15050474602080
UDI-Public(01)15050474602080(17)230529(10)60242584
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K160236
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2023
Device Model NumberOPO73
Device Catalogue NumberOPO73
Device Lot Number60242584
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/16/2020
Supplement Date Received by Manufacturer01/20/2021
02/23/2021
Initial Report FDA Received Date01/13/2021
Supplement Report FDA Received Date02/19/2021
03/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-