• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number ROTAFLOW
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/04/2021
Event Type  malfunction  
Manufacturer Narrative
A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
It was reported that the hospital staff accidentally switched off the rotaflow.No patient harm occurred.Complaint id: (b)(4).
 
Manufacturer Narrative
The initial failure description was that the rotaflow was "accidentally shut off by nurse".The device was directly involved in the event.According to the communication grid dated on 2021-01-18 a getinge service technician checked the rotaflow device.The rotaflow is working normally and no parts have been exchanged.The device has been returned to clinical use.The failure can be confirmed.The most probable root cause could be determined as user error.The product in question was produced in 2018-03-29.The review of the non-conformities during the period of 2018-03-29 to 2021-01-14 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences can be excluded.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.
 
Event Description
Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key11169013
MDR Text Key227109917
Report Number8010762-2021-00013
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW
Device Catalogue Number70104.6405
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/05/2021
Initial Date FDA Received01/14/2021
Supplement Dates Manufacturer Received01/25/2021
Supplement Dates FDA Received01/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-