| Model Number |
SB6AL20 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During a procedure promoted by stemi, an attempt was made to use on sherpa guide catheter to treat a lesion.The device was inspected and prepped per ifu with no issues noted.Excessive force was used and the device was excessively torqued.It was reported that the catheter was over torqued and kinked while in the patient.The kink was so severe, the catheter would not pass back through non-medtronic sheath and the guide catheter had to be removed.The device was replaced with a 7f system for increased support.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: received for analysis was one sherpa guide catheter.A torsional kink was noted on the shaft of the catheter distal to the strain relief.Wire braiding was exposed at this torsional kink site.Kinks were noted on the shaft of the catheter distal to the strain relief.No damage was noted to the distal tip.The id of the catheter met specification.The shaft od met specification.No other damage was noted to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|