| Model Number |
0684-00-0470-01C |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during intra-aortic balloon (iab) therapy, the optical fiber pressure signal is sometimes not available.The iab was replaced with a new one.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.Reference complaint # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during intra-aortic balloon (iab) therapy, the optical fiber pressure signal is sometimes not available.The iab was replaced with a new one.There was no patient harm or adverse event reported.
|
| |
|
Event or Problem Description
|
|
It was reported that during intra-aortic balloon (iab) therapy, the optical fiber pressure signal is sometimes not available.The iab was replaced with a new one.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane completely unfolded and blood on the exterior of the catheter and between the catheter and the sheath.The returned sheath was over the catheter and partially covering the rear portion of the balloon.Three kinks were found on the catheter tubing and inner lumen approximately 39.1cm, 74.9cm and 76.2cm from the iab tip.A sensor output test was performed and the sensor was found to be within specification.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.The reported event cannot be confirmed by the evaluation.A device and lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint # (b)(4).
|
| |
|