• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD SENSATION 7FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD SENSATION 7FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0470-01C
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.(b)(4).
 
Event or Problem Description
It was reported that during intra-aortic balloon (iab) therapy, the optical fiber pressure signal is sometimes not available.The iab was replaced with a new one.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.Reference complaint # (b)(4).
 
Event or Problem Description
It was reported that during intra-aortic balloon (iab) therapy, the optical fiber pressure signal is sometimes not available.The iab was replaced with a new one.There was no patient harm or adverse event reported.
 
Event or Problem Description
It was reported that during intra-aortic balloon (iab) therapy, the optical fiber pressure signal is sometimes not available.The iab was replaced with a new one.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
The product was returned with the membrane completely unfolded and blood on the exterior of the catheter and between the catheter and the sheath.The returned sheath was over the catheter and partially covering the rear portion of the balloon.Three kinks were found on the catheter tubing and inner lumen approximately 39.1cm, 74.9cm and 76.2cm from the iab tip.A sensor output test was performed and the sensor was found to be within specification.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.The reported event cannot be confirmed by the evaluation.A device and lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION 7FR. 40CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
MDR Report Key11169713
Report Number2248146-2021-00022
Device Sequence Number19827754
Product Code DSP
UDI-Device Identifier10607567109374
UDI-Public10607567109374
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K063525
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2023
Device Model Number0684-00-0470-01C
Device Catalogue Number0684-00-0434
Device Lot Number3000120479
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Initial Date Received by Manufacturer 12/25/2020
Supplement Date Received by Manufacturer01/12/2021
02/10/2021
Initial Report FDA Received Date01/14/2021
Supplement Report FDA Received Date02/05/2021
02/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
Patient Weight55
-
-