| Model Number |
72200419 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure, when the package was opened, the accu-pass could not roll the monofilament out of device.The procedure was completed with the same device with no patient injury or delay.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection revealed that two nylon monofilaments were returned with the device.The needle did not appear to be deformed.The monofilament was not in device and doesn't appear to be used.No debris on device or sutures.Both sutures appeared to be cut to manufacturer intent.A functional evaluation revealed the suture shuttle moves freely with ease.An attempt to push each suture through the shuttle was successful.No binding of any kind.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|