| Model Number |
80440 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly a result of an escape of wbc from the lrs chamber near the end of the platelet collection.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related or that excessive access pressure alerts may have contributed.Investigation is in process.A follow up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: w091020456151 the platelet collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in h.6 and h.10.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.The donor pre procedure cbc results were provided to aid in the investigation.Root cause: the signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly a result of an escape of wbc from the lrs chamber near the end of the platelet collection.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related or that excessive access pressure alerts may have contributed.
|
| |
|
Search Alerts/Recalls
|