| Catalog Number |
66801281 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that immediately after changing the dressing and resuming treatment, the message of blockage alarm was displayed on the screen.The problem was not solved by exchanging the device.There was no delay.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, was used in treatment was not returned for evaluation with all additional information provided we have not been able to establish a relationship between the device and the reported event or determine a root cause.The manufacturing records show no evidence that the product did not meet the specification at the time of manufacture.A complaint history review found other related failures.Probable root cause maybe a connection issue or component failure.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
|
| |
|