| Catalog Number |
2C4005 |
| Medical Device Problem Codes |
Filling Problem (1233); No Flow (2991)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported a y-type tur irrigation set would not flow.During patient infusion, the "flow was restricted".There was no patient injury or medical intervention.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
F10/h6: device code: 2991 corrected to 1233.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.Visual inspection was performed using the naked eye, did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed, including pressure testing and clear passage test.And it was noted, that the flow was restricted, due to the tubing and bushing was crushed.The reported condition was verified.The cause of the condition was, due to improper assembly, during manufacturing process.A batch review was conducted, and there were no deviations found related to this reported condition, during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|