| Model Number |
8888145252 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during washing with a 10 cc syringe and saline solution according to the protocol, the blue luer adapter and tego detached from the catheter extension lumen generating blood loss/leak of less than 10 cc from the catheter lumen extension and connector, the lumen was immediately clamped.Cleaning with chlorhexidine was performed and the catheter lumen was sealed with a sterile 1cc syringe, covered with gauze and fixed to the skin.Saline solution was also used as a cleaning agent on the device.There were no other deficient products used during the event.The catheter was not repaired.There was no medical or surgical intervention needed to prevent a permanent impairment of a function.The event did not to extended patient hospitalization.The physician was informed, and the vascular surgeon assessed the patient for a catheter change.The procedure was completed without problem.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.The returned photo showed the venous luer adaptor was disconnected from extension tube of the catheter.It was reported that the luer adapter detached from the extension tube.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|