|
The following fields were updated: g3, g6, h2, h3, h6, h10.This supplemental report is being submitted to provide the results of the manufacturer¿s investigation.The device evaluation, a review of the device history record (dhr), and a review of the instructions for use (ifu) were conducted during this investigation.The device evaluation found the device was broken in multiple pieces at the distal tip.The rest of the guidewire had debris indicating use but no further damage.The hydrophilic distal tip was broken in three separate pieces.The tip was broken about halfway up the tip and small piece of the coating had come off as well.The inner wire was broken as well.The damage of the tip was consistent with what was reported.The review of the dhr did not find any abnormalities or anomalies identified during production.The device met all specifications upon release.The root cause was unable to be confirmed.It is possible the device was bent sharply or bent too much causing this type of damage.The ifu contains the following statements regarding this event: -"the tips of some metal instruments may cause the coating material on the guidewire to be scraped off under conditions of sharp bending or kinking.If this occurs, it is recommended that any fragments of the outer coating material be removed.While advancing or withdrawing any coated guidewire, avoid sharply bending or kinking the portion of the guidewire that extends beyond the instrument." olympus will continue to monitor field performance of this device.
|