SMITH & NEPHEW, INC. 12.5MM ENTRY REAMER; PROSTH, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Model Number |
71631116 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results of investigation: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The tip of the device is fractured, the fractured piece was not returned, rendering the device inoperable.The device was manufactured in 2008 and shows signs of extensive use.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Internal complaint reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the 12.5mm entry reamer was found defective.No case involved.Results of investigation confirmed that the tip of the device is fractured.
|
| |
|