• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM Back to Search Results
Model Number 22438-19
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/30/2020
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record and complaint history of the reported lot could not be conducted because the lot number was not provided.It should be noted that the emboshield nav6 embolic protection system electronic instructions for use states: the emboshield nav6 device can only be used with the barewire filter delivery wire.Use of the device with any guide wire other than the barewire filter delivery wire will lead to loss of the filtration element during the procedure or an inability to retrieve the filtration element.The investigation determined the reported difficulties of detachment of the filter appears to be related to the use error; however, the reported treatments appear to be related to the operational context of the procedure as a snare device was used to retrieve the filtration element from the anatomy.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that the procedure was performed to treat a lesion in the superficial femoral artery.An emboshield nav6 embolic protection system (eps) was used with an unspecified wire rather than the provided barewire.During retrieval, the filter came off of the wire, so a snare was used to successfully retrieve the device.No portion of the device remains in the anatomy.There were no adverse patient sequelae and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key11174548
MDR Text Key226999103
Report Number2024168-2021-00461
Device Sequence Number1
Product Code NTE
UDI-Device Identifier08717648137754
UDI-Public08717648137754
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number22438-19
Device Catalogue Number22438-19
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/30/2020
Initial Date FDA Received01/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-