|
Additional information: g3, g6, h2, h3, h6 h10.The reported event of a leak in the junction between the graft and valve could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.While the cause of the event could not be established, the warnings section of the instructions for use artmt100104849 rev.A, contains multiple items for guidance to mitigate potential for damage of the valve graft." do not use cutting edge or cutting point needles in the sewing cuff or valve graft.Use of standard round or taper point needles is recommended to prevent cutting the polyester fibers." "avoid excessive manipulation of the sjm masters hp valve graft.If surgical technique requires cross-clamping of the graft, cover the jaws of the clamp with soft tubing to prevent damage to the valve graft." "gently extend the graft lengthwise until the crimps are flat.Avoid excessive tension on the sjm¿ masters hp valve graft." "gently extend the graft lengthwise until the crimps are flat.Avoid excessive tension on the sjm¿ masters hp valve graft." "if a needle is used to de-air the sjm masters hp valve graft, use the smallest possible needle.19 gauge needles are normally sufficient.Use of larger needles or hypodermic needles may result in blood leakage, and repair by suturing may be required." needle is used to de-air the sjm masters hp valve graft, use the smallest possible needle.19 gauge needles are normally sufficient.Use of larger needles or hypodermic needles may result in blood leakage, and repair by suturing may be required.".
|