| Model Number |
0935280 |
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a hair inside the asepto syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed cause unknown.Visual evaluation of the returned sample noted one opened (without original packaging), unused irrigation bulb syringe.Visual inspection of the sample noted that there was a hair on the barrel.This does not meet the specification as "no hair is permitted on the product." a potential root cause for this failure could be defective / contaminated components from supplier¿.The lot number was unknown; therefore, the device history record could not be reviewed.Labeling review was not performed because labeling could not have prevented the reported failure.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that there was hair inside of the asepto syringe.
|
| |
|