• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HALOGEN LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HALOGEN LIGHT SOURCE Back to Search Results
Model Number CLH-SC
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 07/21/2020
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the repair center confirmed the customer's complaint.Based on the results of the investigation, it is likely the event was caused by the user hitting another device on the power switch or it was caused by deterioration over time due to the repeated use of the device.
 
Event Description
The customer contacted olympus to report the power switch was broken and the device had exposed terminals.The device malfunction was found during maintenance and there was no patient involvement.
 
Manufacturer Narrative
This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.Correction to g3 of the initial medwatch.The aware date should be 21-jul-2020.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11178165
MDR Text Key227134070
Report Number8010047-2021-01480
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170024467
UDI-Public04953170024467
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLH-SC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received01/15/2021
Supplement Dates Manufacturer Received10/02/2022
Supplement Dates FDA Received10/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/05/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-