| Model Number |
03P6832 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No additional patient information is available.A review of tickets was performed for lot number 01965un20.The ticket search determined that there is normal complaint activity for the reagent lot.A review of tracking and trending did not identify any trends for the complaint issue.Return testing was not completed as returns were not available.The historical performance of reagent lot 01965un20 was evaluated using field data.The patient data was analyzed and compared to an established control limit.This evaluation indicated that the patient median result for lot 01965un20 is within the established control limits.A review of the manufacturing documentation did not identify any issues associated with the complaint issue.A review of labeling concluded that the issue is sufficiently addressed.Based on the investigation no systemic issue or deficiency of the magnesium assay for lot 01965un20 was identified.
|
| |
|
Event or Problem Description
|
|
The customer generated falsely elevated magnesium results while using the architect c16000 processing module.The customer's normal range is 1.6 - 2.6 mg/dl.The following information was provided: sid (b)(6) nitial result: 7.2 mg/dl, repeat 1.9 mg/dl on a different analyzer.Initial result: 7.2 mg/dl, repeat 1.9 mg/dl on a different analyzer.
|
| |
|