It was reported that this electrode and subcutaneous implantable cardioverter defibrillator (s-icd) exhibited oversensing of noise three hours post implant.Three inappropriate shocks were delivered in two separate episodes as a result.Review of the stored episode noted the electrogram (egm) signal became very small and then a wandering baseline was noted.The noise was able to be reproduced by tapping on the sternal incision and was felt to be related to air entrapment.The primary sensing vector was reprogrammed, the device was programmed off and the physician monitored the patient in the hospital overnight.No further noise was observed the following day.The air was felt to have dissipated and the patient was discharged from the hospital.No additional adverse patient effects were reported.This product remains in service.
|