| Model Number |
213041-LVIS-D-PMA |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformance associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and the delivery system was not returned to the manufacturer for analysis.Intraprocedural or post-procedural images were not provided for review; therefore, the alleged product issue cannot be confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that after deployment of an lvis in the internal carotid artery (ica) to treat a ruptured aneurysm, the proximal markers of the stent did not open.The markers could not be opened with the catheters.No additional intervention was performed.Final angiography demonstrated a mostly patent ica and the aneurysm was successfully coiled.There was no reported patient injury or sequela.
|
| |
|
Search Alerts/Recalls
|