WARSAW ORTHOPEDICS ELEVATE SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
| Model Number |
8880923 |
| Medical Device Problem Codes |
Mechanics Altered (2984); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare professional via field representative regarding patient with spondylolytic spondylolisthesis involved in 5/splif procedure.Levels implanted- l5/s.It was reported that event occurred during surgery, when the cage was stuffed with bone.After filling the cage with bone for insertion into the patient, it was confirmed that the cage did not dilate/expand when it was checked to see if it could be dilated.Hence, cage was not implanted in the patient.The product was replaced with a new cage.Event was associated with the patient.Reported, it seemed that there was a piece of bone in the screw part, when they tried to remove the bone from this cage after surgery, it seemed that the bone was stuck in the screw.There was no delay in overall procedure time.Patient was not hospitalized prolong as a result of the event.No health damage in the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluated summary- visual and microscopic inspection confirmed the peek portion of the spacer has broken away from the s pacer.There are witness marks on the backside of the spacer indicating that the spacer was impacted.Both of the ears on the peek portion of the spacer have broken off indicating that the implant came in contact with bone and was pushed back, causing the ears to break off and that section of the implant to separate.This type of damage is consistent with excessive force.Additional information- d9, g3, h3, h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|