STRYKER NEUROVASCULAR CORK NEUROFORM ATLAS 3.0MM X 24MM NO TIP- PMA; STENT, INTRACRANIAL NEUROVASCULAR Back to Search Results
Model Number M003UZAS30240
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/22/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is not available to manufacturer.
 
Event or Problem Description
It was reported that the stent (subject device) prematurely partially deployed and was unable to be recaptured during the procedure.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual and functional testing as well as physical analysis cannot be performed.The reported complaint could not be confirmed and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.No additional information was received.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned to the as reported.
 
Event or Problem Description
It was reported that the stent (subject device) prematurely partially deployed and was unable to be recaptured during the procedure.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROFORM ATLAS 3.0MM X 24MM NO TIP- PMA
Common Device Name
STENT, INTRACRANIAL NEUROVASCULAR
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
MDR Report Key11182639
Report Number3008881809-2021-00032
Device Sequence Number14474600
Product Code NJE
UDI-Device Identifier07613327382204
UDI-Public07613327382204
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P180031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2024
Device Model NumberM003UZAS30240
Device Catalogue NumberM003UZAS30240
Device Lot Number21459508
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/22/2020
Supplement Date Received by Manufacturer02/23/2021
Initial Report FDA Received Date01/16/2021
Supplement Report FDA Received Date03/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-