Model Number 11403001
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported defective device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
1st fiber: an angiodynamics tm reported an end user experienced an issue with a nevertouch laser fiber.The customer reported that they received two laser kits that where defective /damage one was kinked very severely and the other was cut.Both kits were opened for the same patient/procedure.It was reported the procedure was completed with a new of the same device.There was no patient harm or injury due to this event.The reported device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.
 
Additional Manufacturer Narrative
Returned for evaluation was a nevertouch fiber.As "received", the fiber shaft was a bent 4.7cm from fiber gripper.The fiber was connected to the tagwrite, it was noted that the fiber has not been fired.The customer's reported complaint description of the fiber fracturing is confirmed.A review of the returned sample shows the shaft of the fiber's glass core appears to be crushed at the bend of the shaft; this could occur when handling.As stated in the ifu "avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a diameter of 16cm." the reported defect cannot be determined as it appears the damage to this kit most likely resulted from being subjected to some unusual handling conditions after the kit was packaged.A lot history records search for the nevertouch kit revealed no quality related issues or manufacturing deficiencies at the time of manufacture.During the manufacturing process, the disposable fiber device receives a 100% inspection and an aql inspection in which the quality of the fiber strip is inspected.Prior to packaging, all components are inspected for damage.Labeling review: the instructions for use, which is supplied to the end user with this catalog number, states: "prior to and during use, avoid damaging the fiber by striking, stressing, or excessive bending.Do not coil the fiber tighter than a diameter of 20cm.Clinical safety and effectiveness data is not available for other fiber tip designs and diameters.Prior to and during use, avoid bending the introducer sheath and dilator as this can cause kinks and damage".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENACURE NEVERTOUCH FIBER
Common Device Name
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
MDR Report Key11182947
Report Number1319211-2021-00005
Device Sequence Number19784569
Product Code GEX
UDI-Device IdentifierH787114030010
UDI-PublicH787114030010
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K171921
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model Number11403001
Device Catalogue Number11403001
Device Lot Number5637104
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2020
Supplement Date Received by Manufacturer12/28/2020
Initial Report FDA Received Date01/17/2021
Supplement Report FDA Received Date01/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-