| Model Number |
56-3620 |
| Medical Device Problem Codes |
Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Synovitis (2094); Unspecified Mental, Emotional or Behavioural Problem (4430); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Study: (b)(6).Clinical adverse event received for elevated cobalt levels, alval, altered mental status.Event is serious and is considered severe.Event is definitely related to device and not related procedure.Date of implant: (b)(6) 2006, date of revision: (b)(6) 2020, (right hip).Revision of head and liner.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b1 (product problem), b5 and h6 (clinical codes).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d10 and h6 (device code).
|
| |
|
Event or Problem Description
|
|
Revision operative notes (b)(6) 2020, indicate the patient received a right total hip revision due to right hip pain and high cobalt levels.Intra-operative findings include yellow tinged effusion, synovitis, trunnion corrosion, back sided metal liner corrosion and metallosis.Metal head and liner were revised.The surgery was completed without indication of complication by the surgeon.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|