MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 S-ROM M HEAD 36MM +6; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 S-ROM M HEAD 36MM +6; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 52-8330
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Synovitis (2094); Unspecified Mental, Emotional or Behavioural Problem (4430); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Study: (b)(6).Clinical adverse event received for elevated cobalt levels, alval, altered mental status.Event is serious and is considered severe.Event is definitely related to device and not related procedure.Date of implant: (b)(6) 2006, date of revision: (b)(6) 2020, (right hip).Revision of head and liner.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Added: b5 and h6 (clinical codes).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Revision operative notes (b)(6) 2020 indicate the patient received a right total hip revision due to right hip pain and high cobalt levels.Intra-operative findings include yellow tinged effusion, synovitis, trunnion corrosion, back sided metal liner corrosion and metallosis.Metal head and liner were revised.The surgery was completed without indication of complication by the surgeon.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a device manufacturing (mre) review will not be performed even when product/lot information is known.(b)(4), it has been determined that, for the mom platform and related allegations an mre is not required.H10 additional narrative: added: g4 corrected: d1, d2a, d2b, d4, g1, h4.Udi (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S-ROM M HEAD 36MM +6
Common Device Name
S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key11186575
Report Number1818910-2021-01347
Device Sequence Number1101762
Product Code LPH
UDI-Device Identifier10603295171485
UDI-Public10603295171485
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K120599
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number52-8330
Device Catalogue Number136533000
Device Lot Number2209270
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/07/2021
Supplement Date Received by Manufacturer01/20/2021
02/16/2021
Initial Report FDA Received Date01/18/2021
Supplement Report FDA Received Date02/04/2021
02/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; PINNACLE 100 ACET CUP 58MM; PINNACLE MTL INS NEUT36IDX58OD; S-ROM*SLEEVE PRX ZTT, 20F-LRG; SROM STM STD 36+12L 15X20; UNKNOWN HIP ACETABULAR LINERS
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight102
-
-